How the standards fit together
Read it from the bottom up. The quality system is the foundation, development happens inside it, the standards supply the detail, and the evidence they produce is what goes to a regulator.
Show what applies to
Market accessWhere the evidence goes
EvidenceWhat a regulator reads
Standards
Process standards
Development processInside the quality system
Quality systemThe foundation
Every connection, in words
The same 42 connections the map draws, as a list.
FDA 510(k) builds on
- ISO 13485The FDA’s quality system rule, 21 CFR Part 820, incorporates ISO 13485 by reference.
- Design ControlsDesign controls apply to Class II and III devices under 21 CFR 820.10(c).
- V&V & TraceabilityThe submission is built from verification and validation evidence.
- ISO 14971The submission includes the risk management file.
- IEC 62304FDA-recognized. A declaration of conformity can stand in for part of the software documentation.
- IEC 62366-1FDA-recognized. Supports the human factors evidence where use-related risk matters.
- IEC 81001-5-1FDA-recognized. Supports the cybersecurity information required for connected devices.
- IEC 60601Electrical safety and EMC test reports, where the device is electrical.
- IEC 61010FDA-recognized. Electrical safety test reports for laboratory and IVD instruments.
- IEC 82304-1FDA-recognized. Product-level safety evidence for standalone software.
CE Mark builds on
- ISO 13485Harmonised standard for the quality system the regulations require.
- ISO 14971Harmonised standard for the risk management required by Annex I.
- V&V & TraceabilityTechnical documentation includes verification and validation results (Annex II).
- IEC 62304Treated as the state of the art for the software life cycle that Annex I requires.
- IEC 81001-5-1Evidence for the information security requirements in Annex I.
- IEC 62366-1Evidence for the Annex I requirements on use error and usability.
- IEC 60601Evidence for the electrical safety requirements in Annex I, for electrical devices.
- IEC 61010Evidence for the electrical safety of IVD instruments under the IVDR.
- IEC 82304-1Evidence at the product level for software that is a device on its own.
Health Canada builds on
- ISO 13485Licences for Class II to IV devices need an MDSAP certificate: an ISO 13485 audit plus Canadian requirements.
- V&V & TraceabilityLicence applications for higher-class devices include safety and effectiveness evidence.
V&V & Traceability builds on
- Design ControlsVerification and validation are design control activities (clauses 7.3.6 and 7.3.7).
- IEC 62304Software verification activities and their records (clauses 5.5 to 5.7).
- ISO 14971Each risk control measure is verified as implemented and as effective.
- IEC 62366-1The summative usability evaluation is part of validation.
- IEC 60601Type test reports from the laboratory are verification evidence.
IEC 62304 builds on
- ISO 14971Requires risk management to ISO 14971 (clause 4.2). The software safety class comes from the risk analysis.
- ISO 13485Assumes a quality management system is in place (clause 4.1).
- Design ControlsSupplies the software-specific detail for each design control activity.
IEC 62366-1 builds on
- ISO 14971Use-related risks are analysed and controlled through the ISO 14971 risk file.
IEC 81001-5-1 builds on
IEC 60601 builds on
- ISO 14971Requires a risk management process to ISO 14971. The laboratory reviews the risk file.
- IEC 62304Clause 14 points to IEC 62304 for the software in programmable systems.
- IEC 62366-1Collateral standard IEC 60601-1-6 points to IEC 62366-1 for usability.
IEC 61010 builds on
- ISO 14971For IVD instruments, its risk assessment is normally carried out within the ISO 14971 file.
IEC 82304-1 builds on
Design Controls builds on
A summary for orientation. It is not regulatory or legal advice, and it is no substitute for the standard or regulation itself. Requirements change: confirm the current edition before relying on anything here. Last reviewed November 2026.